Alat Uji NDT
FITUR BARU RANGER :
DENGAN FITUR TERBARU KINI SMARTPHONE ANDA BISA TERHUBUNG DENGAN SURVEYMETER DIGITAL RANGER
Radiation Alert Ranger
Internal Halogen-quenched, uncompensated GM tube with thin mica window, 1.4-2.0 mg/cm2 areal density. Effective diameter of window is 45 mm (1.77 in.).
mR/hr - .001 (1µR) to 100 CPM - 0 to 350,000 µSv/hr - .01 to 1000 CPS - 0 to 5000 Total Counts- 1 to 9,999,000 counts
Accuracy (Referenced to Cs137) Typically ±15% from factory, ±10% with NIST Source Calibration
Detects Alpha down to 2 MeV. Detects Beta down to .16 MeV; typical detection efficiency at 1 MeV is approx. 25%. Detects Gamma down to 10 KeV through the detector window. 3340 CPM/mR/hr (137Cs). Smallest detectable level for 125I is .02 µCi at contact.
Sulfur (35S), Strontium (90Sr/y), Cesium (137Cs), Phosphorus (32P), Carbon (14C), Iodine (131I), Cobalt (60Co), and Alpha
mR/hr .001 - 50 and CPM 1 - 160,000. Pulsating beeper sounds the alert. Adjustable alert levels are used for mR/hr / CPM, and µSv/hr / CPS. 70db @ 1m. Alarm will sound when in Timer Mode when set alarm threshold is reached.
Meter will hold at OVER RANGE in fields as high as 100 times the maximum reading.
Display updates every 3 seconds. At low background levels, the update is the average for the past 30-second time period. The timed period for the average decreases as the radiation level increases.
Graphic Display with Backlight
Red LED flashes with each radiation event
Internally mounted beeper (can be switched off for silent operation)
USB for use with Free Observer USB Software for PCs
Two AA alkaline batteries (included)
140 X 68 X 33 mm (5.5 X 2.7 X 1.3 in.)
Carrying Case, Xtreme Boot, Stand, Mini-USB Cable, Observer USB Software Download , Certificate of Conformance
Wipe Test Plate, Swipes, NIST Calibration
1 year limited warranty