Produk atau jasa ini sudah tidak dijual, silakan hubungi perusahaan bersangkutan
untuk keterangan lebih lanjut.
- Film Badge Foma ( Surveymeter/ Survey Meter/ Pendose/ Pen Dose/ Dosimeter)
Film Badge Foma ( Surveymeter/ Survey Meter/ Pendose/ Pen Dose/ Dosimeter)
Update Terakhir
:
13 / 01 / 2020
A. Perdagangan Alat-Alat Proteksi Radiasi, : Surveymeter, TLD, Pendose, Shielding Pb, dll
B. Jasa Pengurusan Ijin Alat Radiasi BAPETEN, : Ijin Impor, Pengalihan, Pemanfaatan, Perpanjangan
C. Jasa lain-lain, : Kalibrasi Alat Ukur Radiasi, Sertifikasi Bebas Radiasi dan Kontaminasi, Uji kebocoran, tes Usap, Survey dan Analisis NORM dan TENORM
Detail Film Badge Foma ( Surveymeter/ Survey Meter/ Pendose/ Pen Dose/ Dosimeter)
PERSONAL MONITORING FILM
FILMS FOR PERSONAL DOSIMETRY
General information
Foma PMF is a set of two films intended for personal dosimetry of gamma radiation, x-radiation and electrons:
Foma DF10 is a high speed film double coated on a blue coloured polyethylene terephthalate base.
Foma DF2is a lowspeed ( emergency) film.
Classification: Personal photographic dosemeter ISO 1757 -3-W-2.The films are supplied in the 3 x 4 cm size ( according to ISO 3665) . One package contains 150 pieces of DF10+ DF2 film sets packed in light-proof and moisture-proof packing.
The films are intended for use with dosimetric film holders.
Depending on the processing method they can be used for measurement of personal dose equivalents of photon radiation in
the energy span from 15 keV to 6 MeV in the range from 0.1 mSv to 2 Sv.
Darkroom light
Due to their high speed, the films should be processed under indirect red safelight ( recommended filters Agfa R1 or R4) using
a 25 watt bulb placed at a distance of at minimum 75 cm from the film.
Processing
Dosimetric Foma films can be processed using the powderx-ray
developer Foma DP with Foma DP-R replenisher and rapid fixer
Fomafix P with replenisher. Application of a mildly acidic stop
bath is recommended between developing and fixing.
Film processing is performed at temperatures from 19 to 21 C.
Development time depends on developer concentration,
temperature and on the intensity of solution exchange in the film
vicinity. It must be determined experimentally. Typical values are
between 5 and 7 minutes.
Confectioning
The film is supplied in the 3 x 4 cm size ( according to ISO 3665) .
One package contains 150 pieces of two sheet moisture -proof
packets from plasticized PVC.
Dosimetric characteristics of the films and their optimisation
The basic dosimetric characteristic is the optical density ver sus
ionising radiation dose curve of the film. As opposed to the
visible radiation, this curve is linear with a dose from the lowest
doses. If the films are to be used for personal dosimetry
purposes, the lowest measurable dose should be approximately
0.1 mSv according to international recommendations ICRP 60
and ICRP 75. This can be achieved with the films Foma DF10
experimentally by an appropriate selection of developing
conditions. The procedure is as follows. The set of personal
monitoring films is expo sed to gamma radiation ( emitter Cs -137
or Co-60) with kerma in the air in the range from 0.1 mGy to ca.
40 mGy. These films are then developed together with films that
were not exposed to the radiation ( fog or background) in given
equipment for a varietyof time periods ( approx. from 5 to 7
minutes) .
The determination of the optimum development time is then
based on the condition that for the high speed film DF10 the
optical density of background ( fog) increased by 4 Ã of the
background must correspond wit h kerma in the air lower than 0.1
mGy ( read from the gamma radiation calibration curve) .
The lowest measurable dose is
0.08 mSy ( ± 30 % ) and the
measurement uncertainty is better
than ± 15 % for values higher than approx. 0.2 mSv. The
requirements of the recommendations ICRP 60 and ICRP 75 are
thus met by a considerable margin ( uncertainty from -33 % to
+ 50 % in annual dose limit) . The corresponding curve forx -radiation with energy 49keV ( approx. 70 kV + 0.5 mm Cu) is
shown in fig. 2.
Calibration curves for emergency films Foma DF2 are shown in
fig. 3 and 4 for gamma radiation andx -radiation with maximum
efficiency ( 49 keV) respectively. These curves were obtained
under developing conditions determined for high speed films
Foma DF10 as optimum. It is evident from figs. 1 and 3 that the
ranges of gamma radiation doses overlap with a considerable
margin and the upper limit of the measuring range is above
1 Gy. If the developing time is reduced by 50% , it will be possible
to use Foma DF2 filmsf or measuring the doses up to at least 2
Gy. However, the lower limit of the measuring range will also be
increased.
Energy dependence of Foma films is shown in fig. 5. As a result
of this dependence the speed of DF10 is approximately 17x
higher forx-radiation with energy 49 keV ( ca 70 kV + 0.5 mm
Cu) . This fact is apparent from the comparison of fig. 1 and fig. 2
( or fig. 3 and fig. 4) . For these reasons, in film dosimetry this
energy dependence must be compensated for using some of the
published methods. Filtration analysis method yields very good
results in the whole energy span 15 keV – 6MeV as it is also
shown in fig. 5.
Fading, i.e. decrease in optical density with time after exposure
to radiation, does not exceed 10% during 3 months in the case
offilms Foma DF10 provided that the ambient temperature is not
higher than 30° C and relative humidity does not exceed 60% .
Fading can be compensated for by an appropriate selection of
irradiation time of the calibration film.
Storage
The method of storage is an important condition for the
preservation of the dosimetric parameters of the films.
A temperature of about 10° C and relative humidity not exceeding
60% are recommended. The films must be outside the reach of
the effects of ionising radiation and aggres sive vapours.
Tampilkan Lebih Banyak